Maintenance Procedures
Chapter 4
31
7. Reverse the above steps to install the replacement cable.
8. Test nurse call function for proper operation before returning bed to service.
End of procedure.
9. Identify the right (
F
, fig. 4.5, page 29) or left (
G
, fig. 4.5, page 29) nurse call cable wires, cut
the cable tie if necessary.
10. Loosen the two screws holding the two groups of red and black wires on the multiple
connector (
C
, fig. 4.5, page 29), remove both groups of wires and segregate the two red and
black wires of the defective nurse call control.
11. Remove the four screws (
D
) holding together the two sections of the nurse call control
housing (
E)
.
12. Remove the two screws (
H
) holding the switch support to the housing. Remove the nurse
call button (
J
) and clean it as well as its housing, if dirt has accumulated.
13. Test the switch and the two sections of the nurse call cable to determine which is
responsible for the problem.
Note that in prior versions of the GO
BED
(beds with a serial
number
[
C02025), the nurse call cable is in one section, running from the nurse call switch
to the multiple connector located in the head end casing.
- if the switch is defective but the cable functions properly, proceed with step 14 through
step 18 and end procedure.
- if one of the cable sections is defective, read the following note and end procedure.
N
N
O
O
T
T
E
E
Replacing a defective nurse call cable is a complex task requiring a good knowledge of the
product. We recommend that you contact our Technical Service department (see section 1.2) in
order for them to suggest an appropriate solution.
End of procedure.
14. Unsolder the two nurse call cable wires (
K
) from the defective switch (
L
). Note the position
of each soldered wire on the switch.
15. Unsolder from the replacement switch the two wires of the cable attached to it. Keep the
cable as an on-hand replacement part.
16. Solder to the replacement switch the two wires of the original nurse call cable.
17. Reverse the above steps to install the replacement nurse call switch and finalize installation
of the nurse call assembly.
18. Test nurse call function for proper operation before returning bed to service.