5.2 REMOVAL OF MOTION INTERRUPT PANEL
Required Tools:
5/16” Nut Driver
FOOT END
A
B
Removal Procedure: (This procedure will take approximately 10 minutes)
1. Unplug the power cord from the wall receptacle.
2. Using a 5/16” nut driver, remove the 2 screws (A) holding the interrupt panel (B) to the litter at the foot end
of the bed (as shown in the illustration).
3. There is a ground strap between the motion interrupt panel and the litter frame. Using a 5/16” nut driver,
remove the screw and star washer securing the ground strap to the litter frame. Save the screw and star
washer for reuse.
4. Lift up and forward on the head end of the panel to disengage the hangers from the litter top.
5. Carefully lower the panel and allow it to rest on the base tubes.
6. Reverse the above steps to reattach the panel.