Chapter Seven – Maintenance Procedures – Foot Board
128
7.1 FOOT BOARD (CONTINUED)
FOOT BOARD INTERFACE PLUG REPLACEMENT
Required Tools:
#1 or #2 Phillips Screwdriver
BOTTOM VIEW OF FOOT BOARD
A
B
A
C
D
E
E
Procedure: (This procedure will take approximately 15 minutes)
1. Unplug the bed power cord from the wall socket.
2. Remove the foot board from the bed to access the bottom of the board.
3. Properly ground yourself (see page 10 for static discharge precautions).
4. Using a Phillips screwdriver, remove the screws (D) holding the drawers (E) to the foot board. Pull the
drawers out of the foot board to expose the interface boards.
5. Using a Phillips screwdriver, remove the two screws (A) holding the plug and washers (B) to the foot board.
6. Pull the plug off the board and disconnect the cables at the exposed interface boards. Note proper place-
ment of the cables so they will be reconnected properly.
7. Reverse the above steps to install the new interface plug.
CAUTION
Be sure to install the plug with the flat edge (C) at the top left, as shown in the illustration or the foot board
interface plug will not mate properly with the bed and damage could result.
Summary of Contents for Secure 3000
Page 78: ...Notes 78 ...
Page 106: ...Notes 106 ...
Page 184: ...Notes 167 2 ...
Page 241: ...Appendix B Circuit Board Locations Bed Wiring Diagrams 222 This page intentionally left blank ...
Page 243: ...Appendix B Circuit Board Locations Bed Wiring Diagrams 224 This page intentionally left blank ...
Page 249: ...DH 3 94 3000 000 020 REV B 420 Alcott Street Kalamazoo MI 49001 ...