![Stryker Secure 3000 Maintenance Manual Download Page 103](http://html.mh-extra.com/html/stryker/secure-3000/secure-3000_maintenance-manual_1386019103.webp)
Chapter Five – Maintenance Procedures – Litter
103
5.5 HEAD SECTION (CONTINUED)
HEAD MOTOR REMOVAL AND REPLACEMENT (FOWLER MOTOR KIT P/N 3000–300–703)
Required Tools:
Side Cutters
5/16” Nut Driver
7/16” Socket Wrench
Procedure: (This procedure will take approximately 45 minutes)
WARNING
The head motor is not grounded and may present electrical hazard. Use caution when working around mo-
tors.
1. Run the head section (Fowler) up to approximately 45
for better access to the motor.
2. Unplug the power cord from the wall socket.
3. Remove litter access panels 3 and 4 (see page 94).
4. Remove the motion interrupt panel (see page 95).
5. Using side cutters, cut the cable ties holding the power cords to the litter (to allow clearance for lifting the
motor out of the litter).
6. Using a 5/16” nut driver, remove the two screws (A) holding the CPR cable bracket to the litter and move
the bracket with the attached cables to one side (to allow clearance for lifting the motor out of the litter).
A
Summary of Contents for Secure 3000
Page 78: ...Notes 78 ...
Page 106: ...Notes 106 ...
Page 184: ...Notes 167 2 ...
Page 241: ...Appendix B Circuit Board Locations Bed Wiring Diagrams 222 This page intentionally left blank ...
Page 243: ...Appendix B Circuit Board Locations Bed Wiring Diagrams 224 This page intentionally left blank ...
Page 249: ...DH 3 94 3000 000 020 REV B 420 Alcott Street Kalamazoo MI 49001 ...