![Stryker Secure 3000 Maintenance Manual Download Page 76](http://html.mh-extra.com/html/stryker/secure-3000/secure-3000_maintenance-manual_1386019076.webp)
Chapter Four – Maintenance Procedures – Base
76
4.4 CASTER REPLACEMENT
Required Tools:
Floor Jack
1/2” Socket w/extension
#2 Phillips Screwdriver
Bungee Cord (or equivalent)
Wooden Brace
FOOT END – BOTTOM VIEW
A
Replacement Procedure: (This procedure will take approximately 10 minutes)
1. Unplug the power cord from the wall socket. Remove upper lift cover (see page 67).
2. Be sure the bed’s brakes are on. Using a floor jack and wooden brace, lift the end of the bed with the defec-
tive caster off the floor a few inches. IMPORTANT: See page 69 for proper bed lifting procedure.
3. Using a 1/2” socket with extension under the base, remove the four castle nuts (A) holding the caster as-
sembly to the base.
4. Reverse the above procedure to install the replacement caster.
NOTE
The upper lift cover was raised to allow access to the four carriage bolts which need to be held when the four
castle nuts are reinstalled.