Chapter Four – Maintenance Procedures – Base
88
4.6 LIFT UNIT (CONTINUED)
LIFT MOTOR COUPLER REPLACEMENT (CONTINUED)
C
E
D
F
4. Using a 5/16” nut driver, remove the two screws (C) holding the metal access panel (D) to the lift housing.
5. The motor coupler (E) can now be removed from the lift housing through the access hole.
NOTE
The replacement motor coupler will come with new rubber bushings.
6. Reverse the above steps to install the new motor coupler and bushings.
NOTE
Removal of access cover (F) is shown only for reference in the procedure for lift motor and capacitor removal
and replacement on page 77.