Chapter Seven – Maintenance Procedures – Foot Board
126
7.1 FOOT BOARD
FOOT BOARD HINGE REMOVAL
Required Tools:
#1 Phillips Screwdriver
Procedure: (This procedure will take approximately 10 minutes)
1. Using a Phillips screwdriver, remove the screws (A) holding the door and hinge assembly to the foot board.
2. If replacing the hinge only, use a Phillips screwdriver to remove the screws holding the hinge to the door.
3. Reverse the above steps to attach the replacement door and/or hinge.
A
A
B
C
C
C
FOOT BOARD MODULE REPLACEMENT (SEE PAGE 15 FOR LIST OF FOOT BOARD MOD-
ULE KITS)
Required Tools:
Phillips Screwdriver
Procedure: (This procedure will take approximately 20 minutes)
1. Unplug the bed power cord from the wall socket. Remove the foot board hinge (see above).
NOTE
Regardless of which module is being replaced, the farthest module to the right must be removed first.
2. Pull the module out of the foot board and disconnect the ribbon cable (B) from the module (C).
3. Reverse the above steps to install the new module.
CAUTION
The modules must be overlapped as shown in the illustration or fluids could enter the board cavity.