7.1 FOOT BOARD (CONTINUED)
FOOT BOARD INTERFACE BOARD REPLACEMENT
Required Tools:
Small Phillips Screwdriver
BOTTOM VIEW OF FOOT BOARD
A
B
B
Procedure: (This procedure will take approximately 15 minutes)
1. Unplug the bed power cord from the wall socket.
2. Remove the foot board from the bed to access the bottom of the board.
3. Properly ground yourself (see page 10 for static discharge precautions).
4. Using a Phillips screwdriver, remove the screw (A) holding the drawer (B) to the foot board.
5. Pull the drawer out of the foot board to expose the interface board.
6. Disconnect all cables from the board. Note proper placement of the cables so they will be reconnected
properly. (See page 218 for foot board wiring diagram.)
7. Using a Phillips screwdriver, remove the four screws holding the interface board to the drawer.
8. Reverse the above steps to install the replacement interface board.