6. Reverse steps to reinstall.
7. Calibrate the scale system. See
8. Verify proper operation before you return the product to service.
B
Ba
atttte
erryy rre
ep
plla
acce
em
me
en
ntt,, sscca
alle
e o
op
pttiio
on
n
C
CA
AU
UT
TIIO
ON
N
•
Always use 1.5 VDC alkaline batteries.
•
Always remove the batteries from the product during long periods of disuse.
The
icon will display when the scale batteries need to be replaced.
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
#2 Phillips screwdriver
P
Prro
occe
ed
du
urre
e::
1. Use a #2 Phillips screwdriver to remove the two screws (A) that secure the battery cover (B) to the enclosure (Figure
19).
B
A
F
Fiig
gu
urre
e 1
19
9 –– B
Ba
atttte
erryy rre
ep
plla
acce
em
me
en
ntt
2. Discard the batteries.
3. Reverse steps to reinstall.
4. Calibrate the scale system. See
5. Verify proper operation before you return the product to service.
H
He
ea
ad
d sse
eccttiio
on
n rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
(2) 1/2” combination wrench
P
Prro
occe
ed
du
urre
e::
1. Raise the head section to the highest position.
2. Lower the rails all the way down.
3. Apply the brakes.
EN
32