2. Raise the siderails to the highest position.
3. Lift the foot section and fold it toward the head end of the product.
4. Using a #2 Phillips screwdriver, remove the seven screws that secure the foot cover plate to the litter frame (Figure 16).
Remove and save the cover plate. Save the screws.
5. Using a 9/16” combination wrench, remove the bolt (J), washer (H), and spacer (G) that secure the fixed top socket (B)
to the accessory bracket (Figure 15). Save the bolt, washer, and spacer.
6. Ground yourself. See
Protecting against electrostatic discharge (ESD)
7. Using diagonal pliers, cut the cable ties that secure the cable to the lower plate and the accessory bracket (E) (Figure
15).
8. Lift the accessory bracket cover while you slowly pull on the cables. Turn the accessory bracket cover toward you until
the inner side of the cover is accessible.
9. Using a small regular screwdriver, remove the load cell cable connectors (T) from the control board (Figure 15).
10. Disconnect the remaining cable connected to the control board. Note the position of the cables for reinstallation.
11. Using a #2 Phillips screwdriver, remove the screw that secures the control board grounding wire to the accessory
bracket. Save the screw.
12. Using a #2 Phillips screwdriver, remove the five screws (D) that secure the control board (C) to the accessory bracket
cover (Figure 15). Remove and discard the control board. Save the screws.
13. Reverse steps to reinstall.
N
No
otte
e -- For load cell cable connections to the control board, see Figure 17.
14. Calibrate the scale system. See
15. Verify proper operation before you return the product to service.
L
Lo
oa
ad
d cce
ellll rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
#2 Phillips screwdriver
•
Small regular screwdriver
•
9/16” socket
•
9/16” combination wrench
•
Diagonal pliers
•
Sawhorse
•
0.036 in. shim
•
Floor jack
•
Torque wrench
P
Prro
occe
ed
du
urre
e::
1. Raise the litter and the siderails to the highest position. Lower the endrail on the side of the load cell to be replaced. If it
is a fixed rail, see the
(page 6) to temporarily lower the rail.
2. Lift the foot section and fold it toward the head end of the product.
3. Using a #2 Phillips screwdriver, remove the seven screws (A) that secure the foot cover plate (B) to the litter frame
(Figure 16). Remove and save the cover plate. Save the screws.
1900-009-002 Rev AA.6
29
EN