34
www.stryker.com
EN
0702-002-700 Rev-S
Specifications
Model:
Neptune 2 Rover-ULTRA
REF:
0702-001-000
Electrical Power
Requirements:
120 V
, 60 Hz, 12.0 A, single phase
20 V
, 3 A during docking procedure; rover receives power from docker REF 0702-014-000
Product Safety Certification:
CSA International
Canadian Standards Association (CSA)
CAN/CSA-C22.2 No. 60601-1:08
, Medical Electrical Equipment — Part 1: General Requirements for Basic Safety and
Essential Performance
CAN/CSA-C22.2 No. 601.1-M90
, Medical Electrical Equipment — Part 1: General Requirements for Safety
American National Standards Institute (ANSI)/Association for the Advancement of Medical Instrumentation
(AAMI)
ANSI/AAMI ES60601-1:2005,
Medical Electrical Equipment — Part 1: General Requirements for Basic Safety and
Essential Performance
; Consolidated Reprint (2009); Amendment 2 (2010)
Underwriters Laboratories (UL)
UL 60601-1,
Medical Electrical Equipment, Part 1: General Requirements for Safety — First Edition
; Revisions through
and including April 26, 2006
International Electrotechnical Commission (IEC)
IEC 60601-1:2005,
Medical Electrical Equipment — Part 1: General Requirements for Basic Safety and Essential
Performance
; IEC Corrigendum 1 (2006); IEC Corrigendum 2 (2007)
IEC 60601-1:1988,
Medical Electrical Equipment — Part 1: General Requirements for Safety - Second Edition
;
Amendment 1 (1991); Amendment 2 (1995); Corrigendum 1 (1995)
AS/NZS 4268: 2008
N 17693
Dimensions:
Width:
48.3 cm [19 inch]
Height:
259 cm [102 inch] with IV pole up; 177.8 cm [70 inch] with IV pole down
Depth:
58.4 cm [23 inch]
Mode of Operation:
Continuous
Adjustable Suction Limit:
50 to 480 mm-Hg; measured with all ports closed
Vacuum Measurement
Accuracy:
± 5% of full scale (± 24 mm-Hg)
Suction Limit Accuracy:
< 24 mm-Hg or 10% of setting
0000016914, Rev. S Effective Date: Mar 11, 2016 9:46:10 AM