NOTES:
▪
A maximum of three EMPTY 4L CANISTER cycles are allowed
between docking cycles. The number of EMPTY 4L CANISTER
cycles (PREFILL) remaining is displayed.
▪
The values of the fluid volume display will change accordingly
after the EMPTY 4L CANISTER action is complete.
▪
Liquid prefill is added to the 4L canister after the EMPTY 4L
CANISTER action is complete. The sound of the pump adding
liquid prefill is normal.
▪
The 4-liter and 20-liter canisters will return to their previous
suction limit settings after the EMPTY 4L CANISTER action is
complete.
c. If the 20-liter canister is full, shutdown the rover for relocation.
Dock the rover to empty the full canister(s). See the
To Dock the
Rover
section.
Fluid Volume Indicators
NOTES:
▪
The conditions, indications and actions described in this table
pertain to one or both canisters.
▪
To adjust the settings of the ALMOST FULL audible indicators,
see the
Systems Setup Options
section.
FLUID VOLUME CONDITION INDICATION
ACTION
Almost Full – The fluid
volume level is near full
capacity. Fluid suction will
soon stop in the full canister.
Five medium
beeps; fluid
level display
value flashes
Prepare to switch
to an alternate
suction source.
FULL – The fluid volume
level is at full capacity. Fluid
suction has stopped in the
full canister.
Continuous
beeping; fluid
level display
flashes FULL
Push the button
next to the OK
icon to stop
the continuous
audible beeps.
Switch to an
alternate suction
source.
Relocate and
dock the rover to
dispose of waste.
After the Procedure
To Shutdown the Rover
WARNING:
▪
BLOODBORNE PATHOGEN HAZARD
▪
ALWAYS wear PPE when operating or handling this equipment.
▪
ALWAYS leave tubing attached to the manifold and close
unused ports during disposal.
▪
ALWAYS follow local regulations regarding proper handling and
disposal of biohazard waste.
Failure to comply may cause infection and result in healthcare
staff injury.
1. To Remove Fluid Suction Components
a. With suction active, gather the suction tubing toward the manifold
port to purge the tubing of fluid waste. DO NOT remove any
attached suction tubing from the manifold (Figure 33).
1a
Figure 33 To Gather Suction Tubing
0000016914, Rev. S Effective Date: Mar 11, 2016 9:46:10 AM