24
www.stryker.com
EN
0702-002-700 Rev-S
To Shutdown the Rover (continued)
b. Turn the SUCTION LIMIT SETTING dial to zero for the canister
with the manifold to be removed (Figure 34).
c. Push the SUCTION button on the control panel to stop fluid
suction.
1c
1b
Figure 34 To Shutdown Suction
d. Rotate the manifold until it is unlocked, then pull the manifold and
attached suction tubing out of the receptacle, always keeping the
manifold in a horizontal orientation (Figure 35).
e. Properly dispose of the used manifold and attached suction
tubing always maintaining the manifold in a horizontal orientation.
f. Remove all the remaining fluid suction accessories as required.
1e
1d
Figure 35 To Remove Manifold and Tubing
2. To Remove Smoke Evacuation Components
(Figure 36)
a. Push the EVACUATE SMOKE button on the control panel until
smoke evacuation stops and the indicator is not illuminated.
b. Remove the smoke evacuator tubing with any attachments.
2b
2a
Figure 36 To Remove Smoke Evacuation Components
c. Properly dispose of the used smoke evacuator tubing and any
attachments.
3. To Remove IV Pole Components
(Figure 37)
a. Push the IV POLE DOWN button to lower the IV pole.
b. Remove any irrigation bags on the IV pole as required.
3a
3b
Figure 37 To Remove IV Pole Components
0000016914, Rev. S Effective Date: Mar 11, 2016 9:46:10 AM
Printed Date: Feb 24, 2020 18:26:59 GMT