To Operate the Rover (continued)
b. From the smoke evacuation information area of the CONTROL
screen, push the buttons next to the arrow icons to increase or
decrease the smoke evacuator power setting as required
(Figure 28).
2b
SMOKE
20%
AUTO
Figure 28 To Adjust Smoke Evacuation
3. To Adjust the IV Pole
Push and hold the IV pole UP or DOWN ARROW buttons to raise or
lower the pole height, respectively (Figure 29).
3
Figure 29 To Adjust IV Pole
4. To Manage Full Canisters
WARNING:
Make sure suction is not immediately required
before pressing the EMPTY 4L CANISTER button. Suction
will be lost in both canisters for at least 30 seconds after
the button is pressed.
NOTE:
See the
Fluid Volume Indicators
table for conditions,
indications and actions related to a full or almost full canister.
a. If the 4-liter canister is full, push the EMPTY 4L CANISTER button
on the control panel (Figure 30).
Figure 30 To Empty 4-Liter Canister
NOTE:
If the 20-liter canister has insufficient capacity to receive the
4-liter canister fluid, a screen will appear indicating the rover will
require shutdown and docking to empty its full canisters. Push the
button next to the ESC icon to acknowledge the need to dock the
rover (Figure 31).
SMALL CANISTER
VOLUME WILL NOT FIT
IN LARGE – DOCK
ROVER WHEN ABLE
ESC
NOTE
Figure 31 To Acknowledge Need to Dock Rover
b. If the 20-liter canister can receive the contents of the 4-liter
canister, a WARNING screen will appear. Push the button next
to the OK icon if the loss of suction will not compromise patient
safety (Figure 32).
SUCTION WILL BE
LOST TEMPORARILY –
PREFILL
REMAINING
3
OK
ESC
WARNING
Figure 32 To Confirm Empty 4-Liter Canister