SYMBOL
DEFINITION
PROTECTIVE EARTH (GROUND)
MANIFOLD PORT
FLUID SUCTION FILTER
COMPARTMENT
SMOKE FILTER COMPARTMENT
AUXILIARY SUCTION PORTS
Suction Limit Setting Dial Colors
NOTE:
Each color is associated with a suction range as shown in
the suction chart located below the control panel. See the
Features
section for more information.
COLOR
SUCTION LIMIT
RANGE
INCREMENTS
Orange –
MAX(IMUM)
480 mm-Hg
Not Applicable
HIGH
120 to 480
mm-Hg
20 mm-Hg
Yellow –
MED(IUM)
80 to 115
mm-Hg
5 mm-Hg
Green –
LOW
50 to 75
mm-Hg
5 mm-Hg
White –
OFF
0 mm-Hg
Not Applicable
Fluid Level Display Screen Messages
MESSAGE
DEFINITION
FULL
Indicates a canister is full (Figure 10). See the
Fluid Volume Indicators
table.
Error
Indicates a serious error has occurred. See the
control panel display for the specific error message
and the
Error Code Messages
section.
-----
Dashes indicate the value of the volume is in
transition or an error has occurred. The system
may still be used.
FULL
mL
FULL
mL
Figure 10 Fluid Level Display
Audible Event Indicators
This table describes the audible indicators associated with each type
of system event.
TYPE
INDICATION
EXAMPLE
Button Actuation
One short beep
EMPTY 4L CANISTER
button is pushed; EMPTY
4L CANISTER is started.
RESET VOLUME button is
pushed.
Task Completion Three short beeps
Docking cycle is
complete.
4-liter canister has been
emptied into the 20-liter
canister.
General
Notification
One long beep
Message or error
information appears on
control panel display
screen. See the
Error
Code Messages
section.
Lost
Functionality
Two long beeps
Message or error
information appears on
control panel display
screen. See the
Error
Code Messages
section.
NOTE:
For additional indicators, see the
High Suction Indicator
section and the
Fluid Volume Indicators
section.