Inspection and Maintenance
WARNINGS:
▪
Upon initial receipt and before each use, inspect each component
for damage. DO NOT use any equipment if damage is apparent
or the inspection criteria are not met.
▪
DO NOT disassemble, modify, service, or repair any system
component or accessory, unless otherwise specified. Call Stryker
Neptune Customer Service. See the
Contact Information
section.
NOTES:
▪
Only individuals trained and experienced in the maintenance of
reusable medical devices should install, inspect, and test this
equipment.
▪
For service, contact your Stryker sales representative or call
Stryker Neptune Customer Service. See the
Contact Information
section. Outside the US, contact your nearest Stryker subsidiary.
▪
Maintenance documentation for this equipment is available upon
request to Stryker-authorized service personnel only.
INTERVAL
INSPECTION
CRITERIA
ACTION
Before initial use
Check equipment for
damage or missing
components and for
proper operation.
If damage is
apparent, replace
the equipment.
Make sure the
rover and docker
operate as a system
properly.
See the following
sections:
To Dock
the Rover, To Test
the Rover,
and
To
Adjust the Rover
Settings
.
Before each use and
after each cleaning
Check equipment for
damage or missing
components.
If damage is
apparent, replace
the equipment.
Check for cleaning
induced damage
or unacceptable
deterioration on all
external surfaces
of the rover, such
as corrosion,
discoloration, pitting,
or cracked materials.
Check the
canister(s), smoke
evacuator filter
cover, and infrared
communication
window for cracks or
damage.
INTERVAL
INSPECTION
CRITERIA
ACTION
Before each use and
after each cleaning
Check the two
casters with locks
and make sure
the locks function
properly.
If damage is
apparent, replace
the equipment. See
the
Accessories
section.
Check the power
cord for cuts and the
power cord plug for
bent pins.
Check the power
cord receptacle for
bent pins or bent
contacts.
Check the infrared
communication
window for any dirt
or debris.
Thoroughly
wipe away any
dirt or debris
on the infrared
communication
window. See the
Cleaning
section.
Six months
Check the
replacement date
on the fluid suction
filter label. The fluid
suction filter life is
500 hours.
Replace the fluid
suction filter every
six months or as
indicated on the
control panel display.
See the
Accessories
section and the
instructions for use
supplied with the
filter.
Check the
replacement date on
the smoke evacuator
filter label. The
smoke evacuator
filter life is 80 hours.
Replace the smoke
evacuator filter
every six months or
as indicated on the
control panel display.
See the
Accessories
section and the
instructions for use
supplied with the
filter.
NOTE:
If any component must be discarded, see the
Disposal/
Recycle
section.