CleanStation CSIIP Site Preparation Guide
DOC-03202 Rev. A
9
5 Environmental Requirements
5.1 Temperature and Humidity
The temperature and relative humidity around the CSIIP must be within certain
limits. Peak conditions occur when the CSIIP draws maximum electrical power,
resulting in the heat dissipation of 1010 Watts (3446 BTU/hr).
The environmental control system in the room should ensure that the room
temperature and relative humidity do not exceed the limits listed in the table
below.
Table 4 Room temperature and relative humidity
Range
Temperature
18°C–25°C (64°F–77°F)
Relative humidity
30%–70% (non-condensing), optimal = 55%
If the room temperature and humidity are outside of the specifications
provided above, the Support removal process might be prolonged.
5.2 Ventilation
To ensure optimal air quality around the CSIIP, the room ventilation system
should change the air at least four times every hour.