Technische Beschreibung
Technical description
Descripciones técnicas
Brief description (1 line with approx. 40 characters possible)
Response
Possible cause
102: Position the tubing set again
Remove the tubing cartridge brie
fl
y and then start the pump again.
Last sensor test more than 24 h ago, deviation in pressure values
150: Pump stopped
Device function interrupted, pump will need to be re-activated if error message is
automatically reset.
Communication or pressure measurement disrupted
180: Paused
Transient sensor discrepancy, pump continues automatically
Transient discrepancy between redundant pressure measurement values
190: Cartridge detection error
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Cartridge detection electronic error
191: Cartridge locking error
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Cartridge locking electronic error
200: Hardware initialization error
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Device defect
203: C2B error
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Self-test: Communication error
20C: Restart necessary in < 4 h
After 24 h of continuous operation, the pump can no longer be restarted.
The device has been in continuous operation for more than 20 h.
20D: Sensor test overdue
Turn device off and on again.
The device has been in continuous operation for more than 24 h.
251: C2B error
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
Internal communication error
252: Electronics error
Device function interrupted, pump will need to be re-activated if error message is
automatically reset.
Internal communication error
254: Default restored
Information message, work can continue, if necessary check the time setting
File error
255: Electronics error
Information message, it may be possible to continue work
IO error
256: Date reset
Information message, work can continue
RTC Battery empty
258: USB overcurrent
Information message, work can continue, disconnect USB connection
FE: USB power too high
259: Electronics error
Turn the device off and on again, contact KARL STORZ Service if the error reoccurs.
BE: Internal error
9. 4 Information messages