MC
2
Power Systems User's Guide
Appendix D: System Electromagnetic Emissions and Immunity Declarations
1110227 Revision 2
November 2006
49
Table 206 – Recommended separation distances between portable and mobile RF
communications equipment and the EQUIPMENT or SYSTEM – for EQUIPMENT and
SYSTEMS that are not LIFE-SUPPORTING (see 6.8.3.201 b)
Recommended separation distances between
portable and mobile RF communications equipment and the MC
2
Series
The MC
2
Series is intended for use in an electromagnetic environment in which RF disturbances
are controlled. The customer or the user of the MC
2
Series can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the MC
2
Series as recommended below, according
to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
m
Rated maximum output
power of transmitter
W
150 kHz to 80
MHz
d
= [3.5/
V
1
]
e
P
80 MHz to 800 MHz
d
= [3.5/
E
1
]
e
P
800 MHz to 2.5 GHz
d
= [7/
E
1
]
e
P
0.01 0.12
0.12
0.23
0.1 0.38
0.38
0.73
1 1.2
1.2
2.3
10 3.8
3.8
7.3
100 12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Summary of Contents for MC2 POWER SYSTEMS
Page 1: ...MC2 Power Systems User s Guide November 2006 Document ID 1110227 Revision 2...
Page 12: ......
Page 40: ...MC2 Power Systems User s Guide 1110227 Revision 2 November 2006 28...
Page 44: ...Chapter 7 Troubleshooting MC2 Power Systems User s Guide 1110227 Revision 2 32 November 2006...
Page 50: ...Chapter 10 Battery Disposal 1110227 Revision 2 38 November 2006...
Page 54: ...Appendix B Mechanical Specifications 1110227 Revision 2 42 November 2006...
Page 56: ...Appendix C Power System Specifications 1110227 Revision 2 44 November 2006...