ASTOTHERM PLUS Instructions for Use
12 Maintenance
38
12 Maintenance
ASTOTHERM PLUS does not require preventive maintenance (e.g. replacement
of liquids or components). Recurrent tests shall be performed according to chapter
12.1.
No service or maintenance shall be carried out while in use with a
patient.
WARNING
Risk of injury!
The service personnel must be appropriately trained and qualified.
The ASTOTHERM PLUS does not contain any parts the user can repair.
Therefore, do not attempt to repair the ASTOTHERM PLUS yourself.
Contact your local sales point.
Any repairs (such as, but not limited to, changing the power supply cord) to
the equipment may only be carried out by persons authorised and qualified
by the manufacturer.
Modifications to the device are not permitted.
The accessories specified in
chapter 15 Ordering information, accessories and
may be replaced without restriction by the operating or mainten-
ance personnel.
On request, STIHLER ELECTRONIC GmbH will provide service instructions that
will allow properly trained and qualified persons to repair the parts of the
equipment that the manufacturer has designated as repairable.
Provision of technical documents and/or spare parts is not an authorisation from
the manufacturer to open or repair the equipment.