ASTOTHERM PLUS Instructions for Use
7 Operating states
22
7.4 Increasing/Decreasing the setpoint temperature
Control Panel
Action
1.
Briefly press the “Increase” button
or “Decrease”
button
, while the device is turned on (
On Mode
) or
started (
Heating Mode
) to change the setpoint
temperature within the three indicated temperatures.
2. Confirm the new setpoint temperature within 5 seconds by
pressing the “Start” button
.
Device response
The flashing frame indicates the selected setpoint
temperature
.
“Start” LED
flashes green until to the confirmation with
the “Start” button
.
After confirmation the “Start” LED
lights green and the
frame marks the currently selected setpoint
temperature
.
The displayed actual temperature
rises or falls
according to the selection.
If the newly selected setpoint temperature is not confirmed
within 5 seconds by pressing the
“Start” button
, the
temperature regulation continues working with the initial
setpoint temperature.
During operation, the mean temperature of the heat
exchanger is displayed; this is not equal to the temp-
erature of the medium to be warmed. ASTOTHERM PLUS
neither regulates nor monitors/displays the current
temperature of the medium to be warmed.
7
6
4
10
5
4
5
10
9
4