ASTOTHERM PLUS Instructions for Use
10 Alarms and troubleshooting
34
10.4 Processor alarm
Control panel
Device response
Even when you plug in the power cord ….
the “Alarm” LED
lights yellow.
the acoustic alarm signal sounds every 17 seconds.
none of the buttons can cause a device reaction.
Alarm condition
Program fault.
Possible
reason(s)
►
Required
action(s)
Temporary program fault.
►
Alarm reset (see below).
Permanent program fault caused by defect data file.
►
Return ASTOTHERM PLUS to local sales office.
Possible actions
to clear
1.
Press the “Increase” button
and
the “Decrease” button
at the same time until the device switches to
Standby
Mode
.
2. Unplug the power cord and wait one minute.
3. Plug in again the power cord.
11
7
6
●