10 Alarms and troubleshooting
ASTOTHERM PLUS Instructions for Use
33
10.3 Cable break alarm
Control panel
Device response
The display
shows „- - -„.
The selection frame
flashes.
The green “On” LED
lights.
The “Start” LED
flashes green.
The “Alarm” LED
lights yellow.
The acoustic alarm signal sounds every 17 seconds.
The heating element is switched off.
Alarm condition
A defect of the cable break detection or a temperature
sensor break has been detected.
Possible reason(s)
►
Required
action(s)
Warmer is defective.
►
Return ASTOTHERM PLUS to local sales office.
Possible actions to
clear
Press the “Standby” button
to switch the device to
Standby Mode
.
8
10
3
5
11
2
●