ASTOTHERM PLUS Instructions for Use
12 Maintenance
42
Test 6
Manual excessive temperature cut off 1
Required
Actions
1. Operate the Warmer with the middle setpoint temperature.
2. Hold down the
“Start” button
at least for 3 seconds, to
start the test.
Result
The test is successful, when:
The display alternately indicates
t1
and the actual temperature
and after a short time
-
the
“Start” LED
flashes green,
-
the
“Alarm” LED
lights yellow,
-
the acoustic alarm signal sounds.
The test is not successful if any of the following conditions
occurs:
The display does not indicate
t1
.
The
“Start” LED
does not flash.
The
“Alarm” LED
does not light.
The acoustic alarm does not beep.
For the continuation of the tests, press the
“Start” button
to
switch the Warmer back to the
Heating Mode
.
Test 7
Manual excessive temperature cut off 2
Required
Actions
1. Operate the Warmer with the highest setpoint temperature.
2.
Hold down the “Increase” button
at least for 3 seconds, to
start the test.
Result
The test is successful, when:
The display alternately indicates
t2
and the actual temperature
and after a short time
-
the “Start” LED
flashes green,
-
the “Alarm” LED
lights yellow,
-
the acoustic alarm signal sounds.
The test is not successful if any of the following conditions
occurs:
The display does not indicate
t2
.
The “Start” LED
does not flash.
The “Alarm” LED
does not light.
The acoustic alarm does not beep.
For the continuation of
the tests, press the “Start” button
to
switch the Warmer back to the
Heating Mode
.