3 Important safety information
ASTOTHERM PLUS Instructions for Use
11
CAUTION
Risk of radio interference!
The essential performance can be lost or degrated due to EM disturbances.
As a result, there is the possibility of hypothermia of the patient.
According to the standard IEC/EN 60601-1-2, medical electrical equipment
requires special precautions regarding electromagnetic compatibility (EMC)
and needs to be installed and put into service according the EMC information
provided.
This device/system may cause radio interference or may disrupt the operat-
ion of nearby devices. It may be necessary to take mitigation measures, such
as re-orienting or relocating of ASTOTHERM PLUS or shielding the location.
3.4 Notices
NOTICE
To avoid damage to the warmer:
-
Do not immerse the warmer in liquid.
-
Do not disinfect the warmer with these methods:
steam (autoclave)
hot air
thermo-chemical cleaning solutions.
-
Refer to the specific instructions for use of the disinfectants.
The customer is responsible for the proper packaging and labelling of returns.