Summary of Contents for AMSCO 30 Series

Page 1: ...11002980 Revision C AMSCO 30 Series and AMSCO 50 Series Reprocessing Sink Operator Manual INSTALLATION AND OPERATOR MANUAL ...

Page 2: ...2 11002980_C Reprocessing Sink Operator Manual ...

Page 3: ...s versatile configurations meet the ergonomic needs of a diverse workforce Advisory A summary of the safety precautions to be observed when operating and servicing this equipment follow this section Do not operate or service equipment until becoming familiar with this information Any alteration of this equipment not authorized or performed by STERIS Service Engineering will void the warranty and c...

Page 4: ...4 11002980_C Reprocessing Sink Operator Manual ...

Page 5: ...Safety Precautions i 2 1 Safety Precautions i 3 Installation 3 3 1 Installation Instructions 3 3 2 Pre installation 4 3 3 Installation Procedure 5 4 Operating Instructions 21 4 1 Operating Instructions 21 5 Routine Maintenance 23 5 1 Routine Maintenance 23 6 Troubleshooting 25 6 1 Troubleshooting 25 7 Replacement Parts 27 ...

Page 6: ...vi 11002980_C Reprocessing Sink Operator Manual ...

Page 7: ...igure 3 8 Mount Pure Water Faucet 10 Figure 3 9 Install Lines to Pure Water Faucet 11 Figure 3 10 Mount Water Pistol and Air Gun 11 Figure 3 11 Install Lines to Water Pistol and Air Gun 12 Figure 3 12 Level Unit using Leveling Foot 13 Figure 3 13 Assemble and Adjust Tail Piece to Lever Drain Assembly 14 Figure 3 14 Tail Piece To Lever Drain Assembling 15 Figure 3 15 P Trap configuration and Instal...

Page 8: ...viii 11002980_C Reprocessing Sink Operator Manual ...

Page 9: ...ix Operator Manual Reprocessing Sink 11002980_C LIST OF TABLES Description Page Table 6 1 Check for faulty component s 25 Table 7 1 Replacement Parts and Consumables 27 ...

Page 10: ...x 11002980_C Reprocessing Sink Operator Manual ...

Page 11: ...nexperienced unqualified personnel or installation of unauthorized parts could cause personal injury invalidate the warranty or result in costly equipment damage Contact STERIS regarding service options Use only original parts when servicing the equipment The safety reliability and function of the AMSCO Reprocessing Sink are insured only if STERIS original parts and accessories or parts and access...

Page 12: ...ully raised Unplug unit when cleaning lifting columns or cleaning near controller Causing unit to overheat too many times greatly reduces life of unit and could damage unit making it inoperable requiring replacement Use non abrasive polish to clean the sink Rub in a back and forth motion in the same direction of surface grain Only wipe down sink DO NOT use a sprayer to wash and or rinse sink ...

Page 13: ...Reprocessing Sink 11002980_C Installation 3 1 Installation Instructions F Fiig gu ur re e 3 3 1 1 R Re ep pr ro oc ce es ss si in ng g S Si in nk k Personnel Required Two Personal Protective Equipment Required Safety Goggles or Glasses Gloves ...

Page 14: ... with components should be opened for installation of contents Be sure to remove all hardware before discarding packaging The pre rinse faucet s and DI Deionized Water faucet should be pre assembled using directions from their manufacturer Use caution when handling faucets and or working around sink as the finish can be scratched Verify utility connections are located per equipment drawing specifi...

Page 15: ...nto pre rinse faucet inlets using sealing tape and tighten Prevent damage to threads Avoid cross threading during assembly Insert correct end into the faucet as shown here F Fi ig gu ur re e 3 3 2 2 I In ns st ta al ll l M Mo ou un nt tiin ng g S Sh ha an nk ks s 2 Prepare Support Bracket Remove clamp and set aside F Fiig gu ur re e 3 3 3 3 S Su up pp po or rt t B Br ra ac ck ke et t ...

Page 16: ... Reprocessing Sink Operator Manual 3 Mount Support Bracket Mount support bracket to backsplash using two 20 x machine screws two washer and two 20 nuts F Fiig gu ur re e 3 3 4 4 M Mo ou un nt t S Su up pp po or rt t B Br ra ac ck ke et t ...

Page 17: ... Mount pre rinse faucet to backsplash with mounting shanks through holes shown Secure pre rinse faucet using washers and locknuts on back of unit as shown Reassemble clamp removed in STEP 2 to support bracket F Fi ig gu ur re e 3 3 5 5 M Mo ou un nt t P Pr re e r riin ns se e F Fa au uc ce et t ...

Page 18: ...s shown below F Fi ig gu ur re e 3 3 6 6 I In ns st ta al ll l B Br ra aiid de ed d L Liin ne es s Use thread sealing tape for reducing coupling Ensure all faucet valves are in the closed position to avoid spills on the floor Leave the faucets in the off position until the drain is hooked up Test connections for leaks Any leaks need to be addressed before moving to next step ...

Page 19: ...ned copper NPT mounting shank 11001022 into Pure Water Faucet inlet using sealing tape and tighten Prevent damage to threads Avoid cross threading during assembly Insert correct end into faucet as shown below F Fiig gu ur re e 3 3 7 7 I In ns st ta al lll M Mo ou un nt tiin ng g S Sh ha an nk k o on n P Pu ur re e W Wa at te er r F Fa au uc ce et t ...

Page 20: ...d configuration Two holes are provided on backsplash for either configuration see above Mount faucet through chosen hole and secure using washers and locknuts like STEP 4 and tighten Be sure faucet is plumb Use hole seal 11001552 to fill unused hole or to mount gun tips mounting bracket STEP 18 Insert here to reverse workflow F Fi ig gu ur re e 3 3 8 8 M Mo ou un nt t P Pu ur re e W Wa at te er r ...

Page 21: ...addressed before moving to next step F Fiig gu ur re e 3 3 9 9 I In ns st ta allll L Liin ne es s t to o P Pu ur re e W Wa at te er r F Fa au uc ce et t 9 Mount Water Pistol and Air Gun Mounting flanges are provided on backsplash to connect water pistol and air gun Use thread seal tape on water connection Connect according to workflow or in the desired configuration F Fi ig gu ur re e 3 3 1 10 0 M...

Page 22: ...ne 11001115 for water pistol using brass coupling 11000974 as shown Test connection for leaks F Fiig gu ur re e 3 3 1 11 1 I In ns st ta al lll L Li in ne es s t to o W Wa at te er r P Pi is st to ol l a an nd d A Ai ir r G Gu un n 11 Connect All Supply Lines Connect all supply lines for water and air in accordance with applicable local code standards 12 Locate Unit in Position Move unit into posi...

Page 23: ...vel Unit Using leveling feet located under unit level sink It is ideal to use front side top edges of unit to level accurately Use front and side edges when leveling F Fiig gu ur re e 3 3 1 12 2 L Le ev ve el l U Un ni it t u us si in ng g L Le ev ve el li in ng g F Fo oo ot t ...

Page 24: ...ing plate and ABS adapter To adjust gap Loosen 6 bolts securing straight drain pipe to sink frame Adjust position of straight drain pipe as needed Screw tail piece adapter to lever drain Tighten 6 bolts Tighten stainless nut to adapter not shown CAUTION POSSIBLE EQUIPMENT DAMAGE If gap is too large the tail piece will bottom out as sink is lowered causing damage If gap is too small the tail piece ...

Page 25: ... Configure Install basin drains Components are included to build configuration below or can be used to customize connection Otherparts may be required and should be purchased locally F Fi ig gu ur re e 3 3 1 14 4 T Ta aiil l P Piie ec ce e T To o L Le ev ve er r D Dr ra aiin n A As ss se em mb bl li in ng g ...

Page 26: ...Coupling and a nipple can be moved to opposite end of U piece to lower the drain outlet F Fiig gu ur re e 3 3 1 15 5 P P T Tr ra ap p c co on nf fiig gu ur ra at ti io on n a an nd d I In ns st ta alll le ed d Test drain for leaks 16 Verify Connections Verify all connections water supply air supply faucets water pistol air gun and drains have no leaks Any leaks need to be addressed before moving t...

Page 27: ...for anchors If the seismic anchoring system will not be installed proceed to STEP 17 NOTE Anchors include their own instructions Remove 4 brackets 2 per foot to install anchors then re install brackets and secure Connect Lift Columns to Outlet AMSCO 50 Series Reprocessing Sink Mark location on floor F Fiig gu ur re e 3 3 1 16 6 I In ns st ta al lll S Se ei is sm miic c A An nc ch ho or rs s ...

Page 28: ...wer columns to full retraction lowest setting d Release Down button e Press Down button again and hold 3 5 seconds and then release f This will reset both columns to the controller and ensures they are both at the same height g If this fails to reset the system contact STERIS Service Representative CAUTION POSSIBLE EQUIPMENT DAMAGE HAZARD Causing unit to overheat too many times greatly reduces lif...

Page 29: ... Use hole seal 11001552 to mount bracket to backsplash through unused pure water faucet hole Place rubber seal between back of bracket and face of backsplash Optional NOTE Fasten gun tips mounting bracket 11000525 into wall stud if possible Mount gun tips mounting bracket 11000525 to wall using drywall anchors and screws not provided Place tips from both the water pistol and the air gun into holde...

Page 30: ...20 11002980_C Reprocessing Sink Operator Manual ...

Page 31: ...djju us st tm me en nt t Using the lift for more than 2 minutes could cause heat buildup in controller and result in controller shutting down until the normal operating temperature is reached Controller has overload protection built in EOP and if the unit overheats because of continuous use this feature activates and prevents operation until unit cools down to a safe operating temperature typicall...

Page 32: ...22 11002980_C Reprocessing Sink Operator Manual ...

Page 33: ...l purpose cleaner non abrasive for general stains and or a stainless steel stain remover for stubborn stains Using a damp cloth or sponge apply cleaner in a back and forth motion rubbing in same direction as surface grain Thoroughly wipe off cleaner Sink is to be wiped down only DO NOT use a sprayer to wash and or rinse the sink because of electrical components Use a damp cloth Clean air pistol an...

Page 34: ...24 11002980_C Reprocessing Sink Operator Manual ...

Page 35: ...s stopped then release 2 If initialization is successful a slight up down handshake movement of sink legs occurs 3 Check cable connections Mains cable connected to both control box and power outlet 4 Check for obstructions Check under above and on the sides of sink for any obstructions that could prevent movement in either direction 5 Check for faulty component s Table 6 1 Check for faulty compone...

Page 36: ...26 11002980_C Reprocessing Sink Operator Manual ...

Page 37: ...products proceed as follows 1 Include description and part order number as listed in Table 7 1Table 2 Include quantity for each replacement part requested 3 Include configuration and serial number of the Reprocessing Sink on your order 4 Send your order directly to STERIS NOTE Use only STERIS authorized parts on this equipment Use of unauthorized parts will void the warranty Contact STERIS for rec...

Page 38: ...acement Parts and Consumables Continued Description Part Number Prolystica HP Dye Free Fragrance Free Enzymatic Manual Cleaner 2 5 Gallon 1C2410 Prolystica 2X Concentrate Enzymatic Presoak and Cleaner 2 5 Gallon 1C3310 Prolystica 2X Concentrate Enzymatic Presoak and Cleaner 4x1 Gallon 1C3308 ...