Section 3 — Installation
18
11002980_C
Reprocessing Sink
Operator Manual
18. Test the leg system.
a. Plug power chord for lift system into 120V GFI outlet.
b. Test by raising the lift columns to their full range of height, then lower to original settings.
c. Press Down button to lower columns to full retraction (lowest setting)
d. Release Down button
e. Press Down button again and hold 3-5 seconds and then release.
f. This will reset both columns to the controller and ensures they are both at the same height.
g. If this fails to reset the system, contact STERIS Service Representative.
CAUTION
-
POSSIBLE EQUIPMENT DAMAGE HAZARD:
:
Causing unit to overheat too many times greatly reduces life of unit and could damage unit,
making it inoperable, requiring replacement.