10. For STERIS washers using small 300 mL/minute pump, select the 30 ml/min flow rate to start
your calibration procedure. Select the 120 ml/min flow rate for large 1200 mL/minute pump.
Press Scroll key to select the pump number, press the Up arrow to select the appropriate
flow rate. When all pumps have the correct flow rate press enter.
11. Press Scroll key to select which pump to calibrate or press enter key to skip past calibration
routines.
12. Press the Up key to start the pump speed calibration routine. Press Scroll to select the pump
to be calibrated then press the Up key to start the flow rate test. The pump will delay five
seconds before starting. Make sure to have your graduated cylinder under the discharge tube
before you start the pump. The pump will run for 60 seconds. Capture the fluid volume
dispensed and note total milliliters dispensed.
Summary of Contents for AL-1000
Page 1: ...MAINTENANCE MANUAL STERIS AL 1000 Chemistry Delivery System 12 18 09 P764333 823 ...
Page 28: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 27 Figure 2 ...
Page 29: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 28 Figure 3 ...
Page 30: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 29 Figure 4 ...
Page 31: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 30 Figure 5 ...
Page 32: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 31 Figure 6 ...
Page 35: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 34 Schematic ...