AL
‐
1000
Installation
‐
Operation
‐
Service
Manual
P764333
‐
823
December
18,
2009
22
13.
In
the
next
menu
you
will
increase
or
decrease
motor
speed
to
achieve
30
or
120
ml/minute
flow
rate.
If
the
fluid
volume
dispensed
is
less
than
the
desired
30
or
120
ml/min
increase
the
“New
Speed”
setting
in
increments
of
5%
then
press
enter
to
continue.
Decrease
the
motor
speed
by
the
same
%
if
the
flow
rate
test
results
in
too
much
chemical
dispensed.
Once
you
enter
the
New
Motor
Speed
the
controller
will
show
the
pump
speed
%.
Press
enter
to
continue
calibrating
the
next
pump
and
repeat
steps
10,
11
and
12.
14. If you are running in Signal Mode (for Non-STERIS or HAMO Washers) follow all of the same
procedures for calibration.
Summary of Contents for AL-1000
Page 1: ...MAINTENANCE MANUAL STERIS AL 1000 Chemistry Delivery System 12 18 09 P764333 823 ...
Page 28: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 27 Figure 2 ...
Page 29: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 28 Figure 3 ...
Page 30: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 29 Figure 4 ...
Page 31: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 30 Figure 5 ...
Page 32: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 31 Figure 6 ...
Page 35: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 34 Schematic ...