AL
‐
1000
Installation
‐
Operation
‐
Service
Manual
P764333
‐
823
December
18,
2009
20
7. Press the down arrow key to reset cycle counts to “0000”. Press enter to move forward.
8. Use the scroll key to select Metric volume units. For Prolystica Ultra Concentrates dosing is
programmed in Milliliters not ounces to achieve accurate injections. Standard (U.S. Ounces)
can be used for dilute traditional chemistries. Press enter to move forward.
9. The default operating mode for the AL-1000 is “
Relay
” mode. This operating mode allows the
washer chemical outputs to control when the AL-1000 starts and stops dosing of chemistries.
This mode should be used on most newer STERIS washers. “
Signal
mode should be used
on older STERIS-AMSCO washers, older HAMO washers, and non-STERIS washers.