4. Press Enter key until LOW LEVEL STATE displayed. Press the Scroll key to change LOW
LEVEL SENSOR output from Washer Interface Module from Normally Open or Normally
Closed switch. Most older Reliance washers use NO switch input, newer Reliance washers
such as Vision use NC input. Press enter to move forward.
IMPORTANT NOTE:
Power must be cycled to initiate change from NO or NC state.
5. Press scroll and the down arrow key to reset factory defaults. This not needed on initial
installation.
6. Press the scroll key to choose which pump/chemical is dispensed in each cycle. This will
allow you to count each wash cycle. Pump #2 for detergent is the most common product to
be dispensed in each cycle. Select Pump # 2 press enter.