15. Signal Qualifying time is a signal filtering function that can be used to ignore “ghost signals”
line noise and other voltage related false signals that could cause pumps to run when not
needed. Set a higher Signal Qualifier (factory setting is 3 seconds) to a maximum 9 seconds
if the washer chemical signal causes false triggering of your pumps.
16. In Signal Mode you will program your chemical volumes on this screen. Remember the AL-
1000 in Signal Mode will control the dose volumes (ml’s or ounces).Enter your volume to be
dispensed when the chemical signal from the washer is received. You can also program a
“Delay” time that delays the injection for up to 99 seconds to allow water levels to rise etc.
17.
Self
test
mode
of
operation
is
activated
by
pressing
UP
&
DOWN
arrow
keys
simultaneously.
The
control
will
sequence
turning
pumps
ON/OFF
and
turning
level
sensor
output
to
Washer
Interface
Module
ON/OFF.
COM
ERRORS
displayed
should
stay
at
000;
if
it
counts
up
this
indicates
a
faulty
communication
cable
or
connection.
Press
and
hold
Enter
key
to
resume
normal
operation.
Summary of Contents for AL-1000
Page 1: ...MAINTENANCE MANUAL STERIS AL 1000 Chemistry Delivery System 12 18 09 P764333 823 ...
Page 28: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 27 Figure 2 ...
Page 29: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 28 Figure 3 ...
Page 30: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 29 Figure 4 ...
Page 31: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 30 Figure 5 ...
Page 32: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 31 Figure 6 ...
Page 35: ...AL 1000 Installation Operation Service Manual 764333 823 December 18 2009 34 Schematic ...