Introduction
This section details the DS345's default settings, storing and recalling set-
tings, setting the computer interfaces, and running self-test and autocal.
Default Settings
Pressing [SHIFT][DEFAULTS] recalls the DS345's default settings and clears
any stored arbitrary waveforms. The DS345's default settings are listed be-
low:
INSTRUMENT SETUP
Setting
Default Value
Frequency
1.0 kHz
Arb Sampling Frequency
40.0 MHz
Amplitude
0.01 Vpp
Offset
0.0 V
Inversion
Off
Phase
0.0°
Modulation Enable
Off
Modulation Rate
1.0 kHz
Modulation Type
AM
Modulation Waveform
Sine
Sweep Parameters
1.0 Hz start frequency and start marker ,
100.0 kHz stop frequency and stop marker
AM Parameters
50% depth, sine wave
FM Parameters
1.0 kHz span, sine wave
PM Parameters
45.0° span, sine wave
Burst Parameters
1 cycle
Trigger Source
SINGLE
Trigger Rate
1.0 kHz
Interface
RS232
Baud Rate
9600
GPIB Address
19
Power on Status Clear
On
Storing Setups
To store the DS345's current setup press [STO] followed by a location num-
ber in the range of 0 to 9. Pressing any of the UNITS key to enter the loca-
tion number, the message "Store Done" indicates that all of the settings have
been stored.
Recalling Stored Settings
To recall a stored setting press [RCL] followed by the location number (0 - 9).
Pressing any UNITS key enters the location number, and the message "re-
call done" indicates that the complete settings have been recalled. If nothing
is stored in the selected location or the settings have become corrupted, the
DS345 will display "rcl error".
2-25
Summary of Contents for DS345
Page 2: ......
Page 5: ...DS345 Synthesized Function Generator iii...
Page 20: ...Introduction 2 4...
Page 64: ...Programming Commands 3 14...
Page 72: ...Program Examples 3 22...
Page 78: ...Troubleshooting 4 6...
Page 82: ...Performance Tests 5 4...
Page 101: ...Calibration 6 10...
Page 109: ...Arbitrary Waveform Composer 7 8...
Page 117: ...DS345 Circuitry 8 8...