SSK SDIS - 1050 User Manual Download Page 13

Window –

 measuring window configuration 

For position measurement there is possibility to adjust measuring window -  e.g. narrow max. 
field of view range or adjust start and end limits for analog output signals 4 - 20 mA (menu 
items 

Beg

 and 

End

), and adjust hysteresis to eliminate signal bouncing at the edges of 

measuring window (menu items 

HysB

 and 

HysE

). 

 

Distance –

  measuring window configuration by specifying distances 

There is possibility to adjust measuring window alternative way, by specifing the distance 
between sensor and measured material and distances between start of measured window and 
laser pointer and between laser pointer and the end of window. Note: This configuration is not 
always accurate, because intensity of infrared signal depends on conditions in the place of 
installation.   
 

Outputs –

 analog outputs configuration 

There are two independent analog outputs 4-20 mA in the sensor. It is possible to configure 
each output to represent or position or intensity of infrared signal.  There is also possibility to 
specify value of output in mA, when there is no material detected (menu item 

NoMatOut

). 

 

Filters –

 configuration of filters and averaging 

There is possibility to adjust filters of output signals to improve accuracy or to avoid false 
detection. By adjusting 

SiUp 

and 

SiDo 

values you can specify how many times signal must be 

detected befor output detection signal is set on (or off respectively) . Value is in miliseconds. 
By adjusting 

Pos

 value you can specify moving average of analog output position signal per 

miliseconds. 
 

Tests –

 tests of sensor's function 

There is possibility to test various sensor's functions, e.g. to check if it is installed correctly. 
You can manually set detection signal to on, set analog outputs to specified value in mA, or to 
check detection of infrared signal by switching on infrared LED diode inside the sensor. (Note: 
Sensor's IR-LED is situated near the beginning of sensors field of view, so its function can be 
checked correctly only in provided software, where you can view pulse). All tests are ended 
when you quit menu 

Tests

There is also possibility to check sensor's function remotely via test relay. Detection signal is 
set to on, and analog outputs are set to 12mA. 
 

Comm –

 communication configuration 

There is possibility to set various parameters of RS232 and RS485 serial communication here. 
 

SysInfo –

 information about sensor 

Here you can see model number, serial number, firmware version, uptime from start and actual 
temperature inside the sensor. 
 

Fact.Res

 – reset of all set values to factory defaults. 

 

Logout – 

quit setup mode

 
 

Summary of Contents for SDIS - 1050

Page 1: ...USER MANUAL Universal position sensor and hot metal detector SDIS 1050...

Page 2: ...so that the infrared radiation is reflected to the photocell from each facet of rotating mirror very accurate even in the case when the entire sensing field is covered with hot material Radiation rec...

Page 3: ...ing made of aluminum alloy suport wall with mounting stand adjustable in both directions adapted to the cooling unit with water and blowing air windscreen to prevent pollution connection of water and...

Page 4: ...ndicular to the measured product When selecting the location of the sensor attention to the following must be paid Other source of infrared radiation in addition to the measured material can not inter...

Page 5: ...onnecting the cooling water and air and setting changes the sensor mounting 4 4 1 Cooling The cooling channel housing is positioned allowing the cooling water sensor Conditions for cooling The sensor...

Page 6: ...ron etc The temperature range depends on the infrared radiation of the product Angles standard sensor is supplied with a viewing angle of 50 degrees with a special mask inside the aperture angle can b...

Page 7: ...Operating temperature 10 C to 60 C At an ambient temperature above 40 C it is recommended to cool sensor with industrial water at a temperature up to 25 C pressure of 1 2 bar flow rate of 1 2 l min P...

Page 8: ......

Page 9: ...selected gain Thre 25 selected threshold Mode LS selected operation mode of the sensor LS or HMD Laser 0 press button to light up the targeting laser for 2 minutes Laser 1 Failure message can be flas...

Page 10: ...Set Rel Rel value set is chosen by external relay Man1 measure according ValSet1 Man2 measure according ValSet2 Window Beg 5 beginning of measuring window analog output 4mA position 0 End 95 ending o...

Page 11: ...s setting End 0 90 distance between laser to end of meas window meters Outputs AnOut1 I output signal Analog 1 I intensity P position setting of analog AnOut2 P output signal Analog 2 I intensity P po...

Page 12: ...detection is proportional to material position in the sensor s field of view Additionally sensor can be set to HMD mode hot metal detector sensor only detects if signal from PbS photocell exceeds thr...

Page 13: ...ilters of output signals to improve accuracy or to avoid false detection By adjusting SiUp and SiDo values you can specify how many times signal must be detected befor output detection signal is set o...

Page 14: ...sh pull C gray presence of material detect 24VDC max 50mA H ping neg presence of material detect 0VDC max 50mA M black neg alarm all OK 24VDC max 50mA alarm 0VDC Relay outputs E brown thick w presence...

Reviews: