PERFORMANCE TESTS
5-1
PERFORMANCE VERIFICATION
The performance verification procedures in this section test and verify the performance of the SR715/720 and
compare it to the specifications listed at the beginning of this manual. The first set of tests verify the basic
functionality of the unit. The second set of tests verify the critical specifications of the SR715/720. The results
of each section can be recorded on the test sheet located at the end of this section.
FUNCTIONAL TESTS
These simple tests verify the basic functionality of the SR715/720. They are not intended to verify the accuracy
of the unit.
Necessary Equipment
:
Item
Critical Specifications
Recommended Model
Analog Oscilloscope
100MHz Bandwidth
Tektronix 2445
with x10 10M
Ω
probes
24.9
Ω
Resistor
0.1%
Dale CMF55 or equivalent
402
Ω
Resistor
0.1%
Dale CMF55 or equivalent
6.34 k
Ω
Resistor
0.1%
Dale CMF55 or equivalent
100.0 k
Ω
Resistor
0.1%
Dale CMF55 or equivalent
22 nF Capacitor
1% NPO
muRata Erie RPE series or equivalent
1. Front Panel Test
Verifies the front panel display digits, LED's and keypad.
1) Turn on the unit while holding down the [Display] key. A single segment in the third digit of the left
display should be on.
2) Press the [DOWN ARROW] key to light each segment (seven total) and the decimal for the third and
fourth digits of the left display, for a total of 16 segments. Only one segment or decimal point should
be on at a time. (Pressing the [UP ARROW] will step backward through the pattern.)
3) Press the [DOWN ARROW] key again (17th time) to light all the segments of all 12 digits. The AUTO
Led will also be on.
4) Press the [DOWN ARROW] key repeatedly to light the 25 LED's in the display and the 26 LED's on
the keypad. The LED's turn on one at a time, from top to bottom and left to right, first for the display,
and then for the keypad. Only one LED should be on at a time.
5) After all of the LEDs have been tested, further pressing of the front panel keys will display the key
code associated with each key. Each key has a different key code, starting with 01 at the upper left,
and increasing from top to bottom then left to right.
6) The unit must be switched off to leave this mode.