The display will show:
Press
CALV (F1)
to perform the CAL-V™ verification test.
Note: The MTI10/MTL10 will stop measuring flow when performing this
test.
Press EXIT (F4)
to exit if you do not wish to continue.
To select what the flow output will do during a CAL-V, choose from
these options:
Go To Zero: Flow output will be zero during the test (i.e. 4mA)
Hold Value: Flow will hold last value during the test
Select the option and press
OK (F4)
.
WARNING: If you are using a closed loop control, the system needs to
be taken off-line during the test.
Press
OK (F4)
to start CAL-V™. CAL-V™ test screen:
This test will take up to 4 minutes (less time if there is flow) and will
show the Cal value changing as the power to the sensor is adjusted. The
T=xxx is a count down timer indicating how much time is left to finish the
test. A “Please Wait” message will be flashing on and off on line 2 during
this test.
DIAGNOSTIC
CALV ZRO EXIT
Verifying CAL-V
Cal = 12.7 T=123
Take Control
off-line EXIT OK
Flo: Go to Zero
NXT OK
Verifying CAL-V
Please Wait