Alarm
Code
Reason
Action
32
4 to 20mA for flow rate
is out of range
Refer to the MAIN MENU on p. 56 of this
Manual. Use the Set I/O section to verify
range limits under FLO Set 4 to 20mA.
33
4 to 20mA for
temperature is out of
range
Refer to p. 44 of this Manual. Use the
Set I/O section to verify range limits under
FLO Set 4 to 20mA. Channel #2 can be set
for flow or temperature.
34
Busy
Meter is recalculating new parameters.
35
Sensor Bridge
Shutdown
The MTI10/MTL10 probe is getting too
hot. Open sensor circuit and check sensor
wiring.
36
Database CRC Error
Refer to the PARAMETER MENU 2 section
on p. 42 of this Manual to reset CRC.
Use SPC section of menu to reset CRC.
Contact Spirax Sarco Technical Support for
possible causes.
37
Totalizer Error
Detected
See "Reset the Total and Elapsed Time"
on p. 46 for steps to clear Error Code.
Contact Spirax Sarco for possible causes.
38
CAL-V in progress
Wait until the CAL-V or Cal Set is finished.
39
ADC12 versus ADC24
too far
The tolerance between the 2 ADC is out of
specification. Recalling manufacture default
may correct the problem.
40
CAL-V Diff Fail
The CAL-V Diff Failed. Check sensor wiring
and verify that the sensor's resistance is
correct. Call Tech Support.
41
Zero CAL-CHECK Fail
Allow meter to stabilize for 15 minutes and
perform the test again. If another "Fail" test
results, call Tech Support.
103
Troubleshooting: Alarm Codes
Alarm Codes