32
10. Press ENTER when the conductivity solution value
is correct. “SA” will be displayed to indicate that
the value has been saved. The meter will return to
Normal Mode.
O
C
CON
CAL
O
C
CON
uS
CAL
6.Press CAL key more than 2 seconds
to enter calibration. The probe will
automatically detect the conductivity
solution value and the value will be
flashed on the LCD. ( Fig 21 ) Press
the
△
or
▽
key to adjust the value
on the LCD to match the standard
calibration value.
Fig. 21
7.When the conductivity solution value
on the LCD matches the calibration
solution value, Press ENTER key,
the “SA" will be displayed on the
LCD. (Fig 22).Then the LCD will stop
flashing and return to normal mode
then the conductivity calibration is
completed.
21
Fig 22
NOTE:
The characteristics of the Probe may
degrade with time and usage.
Whenever an E16 occurs after
calibration, please replace a new
probe immediately.
11. Repeat the steps if needed to calibrate for
other ranges.
12. If the probe has degraded over time or through use,
“E16” will display after calibration. Please replace
the probe immediately.
Note…
To exit Calibration Mode without saving the new value,
press CAL/ESC. The meter will retain the previously
saved calibration data for the current range.