30
LCD DISPLAY DESCRIPTION
• The conductivity value is
displayed on the left side of
the LCD.
• The temperature value is
displayed on the upper right
side of the LCD.
• The total number of records saved is displayed on
the lower right side of the LCD.
CONDUCTIVITY PROBE CALIBRATION
The probe should be calibrated before first use.
• Always use a fresh calibration solution because
contaminants in the solution will affect accuracy.
• Calibrate at 2/3 full range under most conditions.
For example, if the measuring range is 0~1999 mS,
use 1413 µS solution to calibrate.
• You will only need to do single-point calibration.
Calibration Procedure
1. Remove the protective cap from the conductivity probe.
2. Rinse the probe by inserting it into de-mineralized or
distilled water for 30 minutes.
3
-Press to confirm the
calibration or parameter
setting.
LCD DISPLAY
1.The pH value will be displayed on the
left side of the LCD
2.The temperature value will be displayed
in the upper right corner of the LCD.
3.The MEMO number which is the total
number of records saved in the meter
will be displayed in the lower right corner
of the LCD
O
C
PH
TEMP
MEMO
ATC
Conductivity probe
4.The conductivity value will be displayed
on the left side of the LCD. The Temp.
value will be displayed on the upper
right corner.
5.The temperature value will be
displayed in the upper right corner
of the LCD.
6.The MEMO number which is the total
number of records saved in the meter
will be displayed in the lower right
corner of the LCD.
O
C
CON
TEMP
MEMO
uS
ATC
LCD DISPLAY
PH probe