17
Conductivity Calibration (View Only) Function (P30)
For use with optional conductivity probe.
This function allows you to view previously stored
conductivity calibration values.
1. Enter Function Mode. Press ▲or ▼ until “CAL P30”
is displayed.
2. Press
ENTER
to view the calibration information for
Range 1. “P32” will be displayed.
CON
CON
uS
PH
%
12
Fig.5
Press
△
or
▽
key in the setup mode to
check the CAL P30 calibration
message.(Fig 6)This feature is only to
facilitate the user to view the previous
conductance calibration message
-Press ENTER key, to enter the
calibration information screen with
P32 . P33. P34. P35 CON. uS or mS
calibration information.(Fig 7)P32 will
display calibration information in
Range1. P33 will display calibration
information in Range 2
.…
P35 will
display calibration information in
Range 4. Press ENTER key to return
to CAL 30 screen.(Fig 6)
P30 Cond. Calibration (Review)
Fig.6
Fig.7
CON
CON
uS
PH
%
12
Fig.5
Press
△
or
▽
key in the setup mode to
check the CAL P30 calibration
message.(Fig 6)This feature is only to
facilitate the user to view the previous
conductance calibration message
-Press ENTER key, to enter the
calibration information screen with
P32 . P33. P34. P35 CON. uS or mS
calibration information.(Fig 7)P32 will
display calibration information in
Range1. P33 will display calibration
information in Range 2
.…
P35 will
display calibration information in
Range 4. Press ENTER key to return
to CAL 30 screen.(Fig 6)
P30 Cond. Calibration (Review)
Fig.6
Fig.7
3. Press
ENTER
to continue scrolling through the
calibration information for Range 2 (P33),
Range 3 (P34) and Range 4 (P35).
4. Press
ENTER
to return to the “CAL P30” display.
5. Press ▲or ▼to continue to the next function or
press
CAL/ESC
to exit Function Mode.