Replace the Battery Contact PCB
1
Remove the chassis. Refer to
2
Remove the connector PCB.
Figure 4-19 Replace the Battery PCBA
3
Remove the three screws holding the battery PCB to the chassis.
4
Disconnect the battery PCB from the CPU.
5
Replace the battery connect PCB.
6
Reinstall the connector PCB.
7
Reinstall the chassis and bezel assembly.
8
Install the battery. Disconnect the AC mains. Make sure the unit
powers ON.