Figure 4-28 Docking Station Bottom View
3
Remove the docking station cover plate. Set it aside.
4
Remove the connection J70 (docking station release button LED
power) from the docking station PCBA. Refer to
.
5
Remove the seven (7) screws from the monitor docking PCBA.
Refer to
Figure 4-29 Docking Station Enclosure
6
To make room to slide out the docking station PCBA, loosen the
six screws on the latch assembly in order. Refer to