Remote Nurse Alert Connector -
14 Pin for option A
The Remote Nurse Alert connector is primarily designed to provide
a contact closure to the outside world to indicate an alarm
condition. The nurse alert output pr12V, return and three
sets of relay contacts. There is one set of relay contacts for each of
the three alarm levels. Each set of relay contacts consists of a
common normally open (NO) and normally closed (NC). Each
contact is able to carry at least 150 mA continuous.
The pinout for this connector is as follows:
Warning:
Data interface connectors on this device are Ground (Earth) referenced!
-
Only connect this device to other medical equipment suitable for
use in the Patient Vicinity.
-
Check Ground (Earth) Leakage after connecting data interface
cables.
Table 2-3 Connector Pinouts
Pin
Alarm Circuit
Meaning
Image
1
Alarm 0 (high priority)
Common
2
Normally Closed
3
Normally Open
4
GND
5
Alarm 1 (medium priority)
Normally Closed
6
Normally Open
7
Common
8
GND
9
+12 V, 140
μ
A
10
GND
11
GND
12
Alarm 2 (low priority)
Normally Open
13
Common
14
Normally Closed
Summary of Contents for qube 91390
Page 1: ...S E R V I C E M A N U A L 070 2451 01 Rev F www spacelabshealthcare com October 2014 91390...
Page 24: ...www spacelabshealthcare com 1 18 91390 Q U BE I NT R O D U C T IO N...
Page 42: ...www spacelabshealthcare com 2 1 8 91390 Q U BE SE TU P...
Page 54: ...www spacelabshealthcare com 3 1 2 91390 Q U BE T H EO R Y...
Page 90: ...www spacelabshealthcare com 4 3 6 91390 Q U BE M AI N T E N A NC E...
Page 120: ...www spacelabshealthcare com 5 3 0 91390 Q U BE TRO U B L E S H O O T IN G...
Page 160: ...www spacelabshealthcare com B 2 0 91390 Q U BE AP P E N D IX B SY M B O LS...