software must be reinstalled. Check that all three files associated
with the software distribution (bootrom.bin, app.bin, and
fd_image.bin) are the same version.
Figure 5-18 Error Log Entry for System Software Failure
Monitor Fails Power-ON Diagnostics
Troubleshooting Method 1: Use the Touchscreen or Mouse
1
Display the
Boot
menu
(
2
Select
D - run diagnostics…
, then
i - Run an individual
diagnostic…
Retest the previously indicated failure.
3
Upon verification of the failure, a message directs you to the field-
replaceable unit that failed.
Troubleshooting Method 2: Use the Remote Terminal—Use When
No Display is Present
Note:
This procedure can only be performed if the serial port is set for
DIAGNOSTICS.
1
The
Serial Diagnostics
is an alternate method to access and
control the
Boot
and
Diagnostics
menus when the display
doesn’t function. Attach a null-modem serial cable between the
serial port on the monitor and a remote terminal. Configure the
remote terminal for 9600 baud, no parity, 8 bits, 1 stop bit, and no
flow control. When set for
Diagnostics
the serial port of the
monitor is fixed at these settings.
2
Power ON the monitor.
Summary of Contents for qube 91390
Page 1: ...S E R V I C E M A N U A L 070 2451 01 Rev F www spacelabshealthcare com October 2014 91390...
Page 24: ...www spacelabshealthcare com 1 18 91390 Q U BE I NT R O D U C T IO N...
Page 42: ...www spacelabshealthcare com 2 1 8 91390 Q U BE SE TU P...
Page 54: ...www spacelabshealthcare com 3 1 2 91390 Q U BE T H EO R Y...
Page 90: ...www spacelabshealthcare com 4 3 6 91390 Q U BE M AI N T E N A NC E...
Page 120: ...www spacelabshealthcare com 5 3 0 91390 Q U BE TRO U B L E S H O O T IN G...
Page 160: ...www spacelabshealthcare com B 2 0 91390 Q U BE AP P E N D IX B SY M B O LS...