www.spacelabshealthcare.com
5 - 2 1
9 1 3 9 0
Q U B E
T
R O U B L E S H O O T I N G
3
Choose
p-print boot parameters
from the
Boot Menu
. Compare
the result with
a
If the boot parameters are incorrect, select
z-zero NV ram
from the
Boot Menu
to reset the boot parameters to their
factory defaults. For changes to take effect, perform a cold
boot. If the system boots properly, verify all Backend
(Biomed/Clinical Menu) configurations.
b
If zeroing the NVRAM does not restore the defaults, replace
the NVRAM or the CPU PCBA.
c
If the boot parameters are correct, choose
a- Run all bootup
diagnostics…
from the
Main Diagnostic Menu
to identify the
source of the failure.
4
If the previous steps do not correct the problem, replace the CPU
PCBA.
High-pitched Tones Heard from the qube Monitor and No
Display is Present
Applicable to units with serial numbers 1390-1xxxxx and higher. The
System Failure Alarm emits a two-frequency high-pitch tone when
one of these four conditions occurred:
•
Serial communications between the main CPU and the Always-
On CPU is interrupted.
•
The main CPU detected a fault and triggers the safety circuit
itself.
•
The Always-On CPU (power supply section) detected a fault and
triggers the safety circuit itself.
•
AC-Mains loss, without battery installed, or battery depleted.
Power off the qube monitor to reset the safety circuit. Access the
error log (Refer to
on page 5-12) to identify the cause of
the alarm.
1
If the error log indicates that the CPU or Always ON CPU caused
the alarm to trigger, replace the CPU PCBA.
2
If the error log indicates that the battery was depleted, (“Vcom
Failed on Battery”) recharge or replace the battery
V3_3V
+/- 6% (3.1 to 3.5 V) Firmware shutdown @ +/- 10% (<3.0 or >3.3)
V3_3AO
+/- 6% (3.1 to 3.5 V)
VM12V
+/-5% (-13.2 to -10.8 V)
BattATemp
Status messages
BattBTemp
Status messages
PCBTemp
Warning at 60
°
C, shutdown at 75
°
C
Description
Notes
Summary of Contents for qube 91390
Page 1: ...S E R V I C E M A N U A L 070 2451 01 Rev F www spacelabshealthcare com October 2014 91390...
Page 24: ...www spacelabshealthcare com 1 18 91390 Q U BE I NT R O D U C T IO N...
Page 42: ...www spacelabshealthcare com 2 1 8 91390 Q U BE SE TU P...
Page 54: ...www spacelabshealthcare com 3 1 2 91390 Q U BE T H EO R Y...
Page 90: ...www spacelabshealthcare com 4 3 6 91390 Q U BE M AI N T E N A NC E...
Page 120: ...www spacelabshealthcare com 5 3 0 91390 Q U BE TRO U B L E S H O O T IN G...
Page 160: ...www spacelabshealthcare com B 2 0 91390 Q U BE AP P E N D IX B SY M B O LS...