www.spacelabshealthcare.com
4 - 5
9 1 3 9 0
Q U B E
M
A I N T E N A N C E
8
Wait for the Reset Monitor dialog box to display.
9
Click (touch) the
Reset Monitor
key.
The monitor resets. After 15 to 20 seconds, a lighted box will
appear in the upper left corner of the screen.
10
Touch the lighted box for one or two seconds until it
disappears.
Three more lighted boxes appear, one at a time, in the upper
right, lower right, and lower left corners.
11
Touch the lighted boxes as they appear.
After you touch these four corners one at a time as the boxes
appear, the touchscreen is calibrated.
Functional Tests
Monitor Self-Test
A power-ON self-test is performed each time the monitor is turned
ON.
•
The power ON/OFF switch illuminates indi5 V is present.
•
If the monitor operates on external AC power, the power LED on
the front bezel also illuminates.
•
If the monitor operates on AC power and batteries are charging,
the battery A and battery B status LEDs on the front bezel flash
ON and OFF at a constant rate. Once batteries fully charge,
these LEDs remain ON. If no battery is present, one or both of
the battery status LEDs remain OFF.
Monitor Functional Tests
This procedure verifies operation of the recorder assembly, network
connectivity, alarm relay output, and external SDLC connections. It
assumes that a 91496-B or 91496-C multi-parameter module (ECG,
RESP, and two PRESS channels) is available for testing purposes. If
these parameters are not available, you can substitute similar
modules.
Verification of Monitor Functions
Note:
To begin the test, make sure that no modules are inserted and no
Flexports or other SDLC devices are connected.
1
Plug in the monitor and verify that the front bezel power LED is
lit. If one or two batteries are installed, verify that the appropriate
battery LED or LEDs are ON solid or flashing.
2
To power ON the monitor, press the power ON/OFF switch.
3
Verify the following sequence of events:
a
The power ON/OFF switch illuminates.
Summary of Contents for qube 91390
Page 1: ...S E R V I C E M A N U A L 070 2451 01 Rev F www spacelabshealthcare com October 2014 91390...
Page 24: ...www spacelabshealthcare com 1 18 91390 Q U BE I NT R O D U C T IO N...
Page 42: ...www spacelabshealthcare com 2 1 8 91390 Q U BE SE TU P...
Page 54: ...www spacelabshealthcare com 3 1 2 91390 Q U BE T H EO R Y...
Page 90: ...www spacelabshealthcare com 4 3 6 91390 Q U BE M AI N T E N A NC E...
Page 120: ...www spacelabshealthcare com 5 3 0 91390 Q U BE TRO U B L E S H O O T IN G...
Page 160: ...www spacelabshealthcare com B 2 0 91390 Q U BE AP P E N D IX B SY M B O LS...