Getting Started
FTS Systems FlexiCool
4
Rev 001, 10/13
© SP Scientific 2013
Setup
The FTS System FlexiCool Immersion Cooler is designed for bench-top installation in
a laboratory environment. Your unit should be installed on a firm, level surface in a
location that is convenient for both operation and service.
NEVER PLACE THE UNIT IN AND AREA WHERE EXCESSIVE HEAT,
MOISTURE OR CORROSIVE MATERIALS ARE PRESENT.
Ambient Conditions
NEVER OPERATE THE FTS SYSTEMS FLEXICOOL IN AN AREA WITH AN
AMBIENT TEMPERATURE ABOVE 31 °C (88 °F).
This unit is designed for operation in a normal indoor environment (systems should
not be mounted outside or otherwise exposed to the elements). The environment
should be free from air containing large amounts of moisture, salt or sulfur.
For best low-temperature operation, consider that the ideal ambient temperature for
your FTS Systems FlexiCool is approximately 22 °C (71.6 °F). Higher ambient
temperatures
will limit the system’s ability to achieve its ultimate low temperature.
Ambient Air Temperature Requirements
12 to 20
C (55 to 69
F)
Acceptable, but expect reduction.
21 to 24
C (70 to 75
F)
Ideal.
25 to 27
C (77 to 80
F)
Acceptable, but a reduction in the cooling capacity of the unit
is to be expected (
i.e.,
about a 1% cooling capacity reduction
per degree above 25 °C (77 °F)).
28 to 30
C (82 to 86
F)
Expect reduced reliability. Warranty may be voided.
Over 31
C (Over 88
F)
Not Acceptable. Warranty will be voided.
Summary of Contents for FTS FLEXICOOL IMMERSION COOLERS FC100
Page 1: ...IMMERSION COOLERS FC50 FC100 OPERATOR S MANUAL FTS SYSTEMS FLEXICOOL IMMERSION COOLERS ...
Page 2: ......
Page 6: ...iv Rev 001 10 13 SP Scientific 2013 ...
Page 16: ...Getting Started FTS Systems FlexiCool 8 Rev 001 10 13 SP Scientific 2013 ...
Page 22: ...Operation FTS Systems FlexiCool 14 Rev 001 10 13 SP Scientific 2013 ...
Page 30: ...Maintenance FTS Systems FlexiCool 22 Rev 001 10 13 SP Scientific 2013 ...
Page 37: ......