Appendix B: Changing Temperature Display
FTS Systems FlexiCool
26
Rev 001, 10/13
© SP Scientific 2013
Use the Down and Up buttons to change the setpoint to -99.9 °F.
10.
Press the SET button to save then press it a second time to exit.
11.
Press Rotate button to access Operation Mode Parameters.
12.
Continue pressing the Rotate button until you access Loc2 (
i.e.,
).
13.
Press the SET button to accept.
14.
Temperature Range
FC100
FC50
Span Limits
(SOLO Input Setup)
(P3-3) Range High (tP-H)
30 °C (86 °F)
30 °C (86 °F)
(P3-4) Range Low (tP-L)
-99.9 °C (-99.9 °F)
-60 °C (76 °F)
Setpoint Recommendation
(Short Spec Temp Control Range)
Do not set above
20 °C (68 °F)
20 °C (68 °F)
Do not set below
-99.5 °C (-95 °F)
-50 °C (-58.0 °F)
Note:
The set value is limited to three digits when in negative (-) range therefore -99.9 is the lowest possible set value.