FTS Systems FlexiCool
Maintenance
Rev 001, 10/13
21
© SP Scientific 2013
Refrigeration System
The FlexiCool refrigeration system compressors are equipped with two safety devices
for protection against low voltage or high temperatures. If the voltage drops below
what is required for the compressor to operate, an overcurrent device turns the
compressor off. If the compressor temperature rises too high due to insufficient
air-cooling, a high-temperature thermostat deactivates the compressor.
Note:
High temperature of the compressor can be caused by blocked airflow or by a
malfunction of the fan motor.
The compressor will automatically start when the condition is corrected, but will turn
off again if the voltage is too low or the temperature is too high.
IF ANY REFRIGERATION SYSTEM PROBLEMS OCCUR MORE THAN ONCE, THE
UNIT SHOULD BE SHUT OFF UNTIL THE CAUSE OF THE PROBLEM CAN BE
DETERMINED AND CORRECTED. ALLOWING THE UNIT TO CONTINUE TO RUN
DESPITE ENCOUTERING A PERFORMANCE PROBLEM MAY LEAD TO FAILURE
OF THE UNIT.