Appendix A: Temperature Calibration
FTS Systems FlexiCool
24
Rev 001, 10/13
© SP Scientific 2013
Push the SET button.
7.
Push the Rotate button until you access the thermocouple probe offset
8.
(
i.e.,
).
Match the temperature displayed on the SOLO controller to the value
9.
displayed on the NIST temperature indicator by adjusting the thermocouple
probe offset. Use the Down and Up buttons to adjust your reading.
Note:
The Down button is used to decrease the value shown on the SV display.
Pressing and holding this button will accelerate the pace at which descending
increments are displayed. The Up button is used to increase the value shown on the
SV display. Pressing and holding this button will accelerate the pace at which
ascending increments are displayed.
Press the Rotate Button until LoC appears in the display (
i.e.,
).
10.
Press either the Down or Up button to access until Loc2 appears (
i.e.,
).
11.
Press the SET button to accept and lock the controller.
12.
Down and Up
buttons